Agilent Technologies N9340A User Manual page 146

Handheld spectrum analyzer
Hide thumbs Also See for N9340A:
Table of Contents

Advertisement

7
Error Messages
Error Code Error Message
162
Cannot open fast
sweep in this
case
159
Error 0x11, DSP
will restart
158
Error 0x10, DSP
will restart
157
Error 0xF
156
Incorrect
alignment file
on flash
153
RF EEPROM
operate fail
152
Alignment file
oversize
151
DSP boot fail
150
Mixer overload
149
Mixer saturate
148
Error 0xE
144
Description
Indicates current settings do not
allow you to open fast sweep.
For example, currently in FFT,
zero span, or any measurement
in power suite does not allow
fast sweep.
Indicates DDC failed. If analyzer
froze, try to restart the analyzer.
Indicates an internal error
occured, wait the analyzer
to reboot. If analyzer froze, try
to restart the analyzer.
Indicates an error in internal
communication.
Indicates an invalid alignment
file.
Indicates an error when writing
alignment file.
Indicates the alignment file has
error or invalid data.
Indicates the analyzer
cannot process
measurement,wait the
analyzer to reboot. If
analyzer froze, try to restart the
analyzer.
Indicates the first mixer in
danger. Either increase input
attenuation or decrease the input
signal level.
Indicates you need to either
increase the input attenuation or
decrease the input signal level.
Indicates the analyzer reads
unreasonable parameter value. If
analyzer froze, try to restart the
analyzer.