6
SCPI Command Reference
• Off
Turns marker off.
Key Access:
Set Center Frequency to the Marker Value
:CALCulate:MARKer
[1]|2|3|4|5|6[:SET]:CENTer
This command sets the center frequency equal to
the specified marker frequency, which moves the
marker to the center of the screen. In delta marker
mode, the center frequency is set to the marker
delta value. This command is not available in zero
span.
Key Access:
Set Reference Level to the Marker Value
:CALCulate:MARKer
[1]|2|3|4|5|6[:SET]:RLEVel
This command sets the reference level to the
specified marker amplitude. In delta marker mode,
the reference level is set to the amplitude
difference between the markers.
Key Access:
Marker On/Off
:CALCulate:MARKer[1]|2|3|4|5|6:STATe
OFF|ON|0|1
:CALCulate:MARKer[1]|2|3|4|5|6:STATe?
This command toggles the selected marker between
on and off.
Key Access:
108
[Marker] > {Normal}
[Marker] > {Delta}
[Marker] > {Off}
[Marker] > {Marker To} > {To Center}
[Marker] > {Marker To} > {To Ref}
[Marker] > {Normal}
[Marker] > {Off}
N9340A User's Guide