• When ON at the completion of each trigger cycle,
the trigger system immediately initiates another
trigger cycle.
• When OFF, the trigger system remains in an
"idle" state until CONTinuous is set to ON or an
:INITiate[:IMMediate] command is received.
On receiving the :INITiate[:IMMediate]
command, it will go through a single trigger
cycle, and then return to the "idle" state.
• The query returns 1 or 0 into the output buffer.
1 is returned when in a continuous measurement
state. 0 is returned when there is only a single
measurement.
*RST State:
Key Access:

Take New Data Acquisitions

:INITiate[:IMMediate]
This command initiates a sweep if not in a
measurement. If in a measurement, it triggers the
measurement. A "measurement" refers to any
function under the MEAS key.
Note that the function is enable only when the
continuous is OFF.
Key Access:
N9340A User's Guide
SCPI Command Reference
Continuous, or On
[BW/SWP] > {Sweep}
[BW/SWP] > {Single Sweep}