6
SCPI Command Reference
Control Limit line Buzzer
:CALCulate:LLINe[1]:BUZZer[:STATe]
OFF|ON|0|1
:CALCulate:LLINe[1]:BUZZer[:STATe]?
This command toggles the audio warning between
on and off. If data exceeds the limit line the buzzer
will sound an alarm.
*RST State:
Key Access:
Control Limit Line Testing
:CALCulate:LLINe[1]:[STATe] OFF|ON|0|1
:CALCulate:LLINe[1]:[STATe]?
This command toggles the limit line testing
between on and off.
If the limit line is turned on, limit pattern is
automatically turned off.
*RST State:
Key Access:
Set Limit Line Y-axis value
:CALCulate:LLINe[1]:Y <value>
:CALCulate:LLINe[1]:Y?
This command sets limit line Y- axis value.
Limit line Y- axis value is set independently without
affected by the X- axis Units.
*RST Value:
Available Unit:
Key Access:
102
Off
[Limit] > {Beep}
Off
[Limit] > {Limit Line}
0 dBm
dBm, -dBm, mV, uV
[Limit] > {Limit Line}
N9340A User's Guide