Preamp

Toggles the internal preamp between On and Off.
Preamp results in a correction being applied to
compensate for the gain of the preamp at 20 dBm
so that amplitude readings show the value at the
input connector.
When the preamp is on, a PA indication appears on the left
N O TE
side of the display.
The preamp is set on in frequency bands from
1 MHz to 3 GHz, otherwise the correction is not
applied even though the PA indication will still
appear.
Key Access: [AMPTD]

Scale/DIV

Sets the logarithmic units per vertical graticule
division on the display.
The Scale/Div function is only available when the Scale Type
N O TE
key is set to Log.
Key Access: [AMPTD]

Scale Type

Toggles the scale type of vertical graticule divisions
between logarithmic and linear.
When scale type sets to Log, a Logarithmic units is
activated ranging from 1 to 10 dB per division.
When Scale Type sets to Lin, the vertical divisions
are linearly scaled and the default amplitude units
are volts. The top of the display is set to the
reference- level value and the bottom graticule
becomes zero volts. (Each division of the tragically
is one- tenth of the reference level in volts.)
Key Access: [AMPTD] > {Scale Type}