Error Code Error Message
147
Incorrect
alignment data
in EEPROM
143
Final IF
overload
138
USB device NOT
ready
136
RBW limit to 30
kHz when in fast
sweep
135
Error 0xB
134
Error 0xA
132
Error 0x8
130
Meas uncal
119
RF Board Changed
116
Cannot
communicate with
RF
112
Error 0x3, DSP
will restart
N9340A User's Guide
Error Messages
Description
Indicates invalid alignment data
occurred.
Indicates you need to either
increase the input attenuation or
decrease the input signal level.
Indicates the USB device is not
detected.
Indicates the analyzer
automatically couple the RBW to
30 kHz when in fast sweep
mode.
Indicates the analyzer encounters
a DLL file open failure. If
analyzer froze, try to restart the
analyzer.
Indicates the analyzer failed at a
dynamic loading process. If
analyzer froze, try to restart the
analyzer.
Indicates some important files
missing.
Indicates the measurement is
uncalibrated due to fast sweeping
through a narrow RBW filter.
Check the sweep time, span, and
bandwidth settings, or use auto
coupling.
Indicates the RF board was
changed, the analyzer needs
re-load the alignment data.
Indicates the MCU cannot find
the RF board.
Indicates an undefined error
occurred when MCU tried to
communicate with DSP. Wait the
analyzer to reboot.