Upgrading Firmware
Make sure at least 10 minutes consistently power supply
CAU T ION
CAU-
when updating firmware. Power cut off during the
updating process can make damage to the instrument.
A N9340A provides an easy and fast access for
firmware upgrade.
1 Setup a folder named N9340DATA in the root
directory of your USB memory stick.
2 Download the firmware update package from
Agilent website (www.agilent.com/find/n9340a)
into the folder named N9340DATA.
3 Insert the USB memory stick into the USB
connector on the top panel of your analyzer.
4 Press [SYS] > {More (2 of 3)} > {Upgrade} >
{Firmware} active the inner updating procedure.
Then the instrument will perform the upgrading
process automatically.
Adding an option
Pressing [SYS] > {More (2 of 3)} > {Option} > {Add}
brings up a dialog box for entering the option
license code. Use the numeric keypad to input the
option license code and use the [Enter] key as a
terminator. If the option license code is identified,
an annotation "Option install successfully" will
appear in the status line, else an annotation
"Invalid option licence" will appear in the status
line.
N9340A User's Guide
Getting Started