Reset/Power-On State - Agilent Technologies 34405A User's And Service Manual

5 1/2 digit multimeter
Hide thumbs Also See for 34405A:
Table of Contents

Advertisement

Reset/Power-On State

The table below summarizes the 34405A's settings as received from the
factory, following power cycling, and following the *RST command received
over the USB remote interface. Non- volatile, user customizable behavioral
differences are shown in BOLD type.
Table 2

Reset/Power-On State

Parameter
Measurement Configuration
Function
Range
Resolution
Temperature Units
Math Operations
Math State, Function
Math Registers
dBm Reference Resistance
Math Register Editing
Trigger Operations
Trigger Source*
System-Related Operations
Power-Down Recall
Stored States
Beeper
Display
Remote/ Local State*
Keyboard*
34405A User's and Service Guide
Factory Setting
DCV
AUTO
5-½ digits
°C
Off, Null
Cleared
600Ω
On
Auto Trigger (Local Mode)
IMMediate (Remote Mode)
Disabled
0-4 cleared
On
On
Local
Unlocked, Local key enabled
Features and Functions
Power-on / Reset State
DCV
AUTO
5-½ digits
User setting
Off, Null
Cleared
User setting
User setting
Auto Trigger (Local Mode)
IMMediate (Remote Mode)
User Setting
No Change
User Setting
On
Local
Unlocked, Local key enabled
2
43

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents