Check Test Weight Calibration; Moisture Curve Calibration; M2.0 Moisture Calibration - Juniper HarvestMaster mirus H2 User Manual

Table of Contents

Advertisement

Mirus for H2 Classic GrainGage

4. 3. 2 Check Test Weight Calibration

Test weight calibration is stable. It rarely changes or loses accuracy. However, because the consequences of
poor calibration can be devastating to your research data, HarvestMaster highly recommends checking the
test weight calibration every year before the harvest season begins, as well as every day before harvest. This
can be done on the Test Weight Diagnostics screen.

4 .4 Moisture Curve Calibration

!
CAUTION: Do not re-wet grain samples. If you need to adjust the moisture content, take a natural
sample and dry it down.
Before harvesting, you need to generate a chamber calibration (moisture curve). Chamber calibrations align
the test weight and moisture reading with a benchtop lab sensor.
For better accuracy, the practices in this guide are designed to streamline the calibration process and teach
users how to create moisture curves.

4.4.1 M2.0 Moisture Calibration

With the release of Mirus 4.3.0 there are two ways to perform a moisture calibration with the same
hardware—the traditional EM moisture curve or the new M2.0 moisture calibration for certain grains. The
new M2.0 moisture model is a preset model that will be offset based on the samples you provide during
calibration. If you are harvesting wheat, corn, barley, oats, canola/rapeseed, or soybean, HarvestMaster
recommends using the M2.0 moisture model. For more information about the M2.0 moisture calibration
see 8.4.1 Instructional Content on page 128.
H2 Classic GrainGage Calibration
1. After calibrating the test chamber, tap the gear icon
below the calibration wizard to open Diagnostics.
2. Hang the calibration weight from the weight hanger
below the test chamber.
3. The Diagnostics screen should report a Chamber
Weight reading that is very close to the value stamped
on the calibration weight. If it's not within .05 lb (23 g)
with the combine off, HarvestMaster recommends a
new weight calibration.
4. Tap the check icon
to close Diagnostics.
51