[3.11] Maintenance & Diagnostics > Test; Overview Of Test; [3.11.1] Test Inputs/Outputs; Testing Of Analog Outputs - Siemens SIPROCESS GA700 Operating Manual

Continuous gas analysis, operating with the local user interface
Hide thumbs Also See for SIPROCESS GA700:
Table of Contents

Advertisement

Functions
6.24 [3.11] Maintenance & Diagnostics > Test
6.24
[3.11] Maintenance & Diagnostics > Test
6.24.1

Overview of test

You can use the submenus in this menu to test inputs/outputs, the display, and the keyboard
for functionality.
Note
"Simulation or substitute value" operating mode (Namur: "Check function")
Activation of a test requires registration at the device. When you activate a test, the device
goes into "Simulation or substitute value" operating mode. The device is in "Function check"
status, but outputs a simulation value or substitute value. The corresponding function control
indicator is displayed with the following symbols:
The function control indicator is hidden once the test is complete. Information on the used
symbols, especially when changing the operating mode, is available in: → Table 7-1 Function
check indicators (Page 255).
See also
LUI symbol sets (Page 255)
6.24.2

[3.11.1] Test inputs/outputs

6.24.2.1

Testing of analog outputs

Requirement
Note
Availability of the menu
This menu is only available if your device has at least one of the option modules (2.1 and/ or
2.2) installed.
● Make sure that a suitable test device is connected.
236
→ SIMATIC PCS7 (symbol "Hand, 2 points")
→ NAMUR
Operating with the Local User Interface
Operating Manual, 06/2017, A5E31930478-05