Alarm, error, and system messages
7.2 OXYMAT 7 message list
Message
No.
Text
11
Maximum temperature exceeded.
Heaters turned off automatically.
15
Sample gas pressure outside
tolerance
18
Reference gas pressure too low
22
Calibration tolerance exceeded at
zero point
23
Validation tolerance exceeded at
zero point
24
Calibration tolerance exceeded at
span point
25
Validation tolerance exceeded at
span point
27
Signal frequency changed
28
Control reserve used up
29 to
Zero drift MR1 to MR4 outside
32
tolerance
33
Measured value greater than full-
scale value
34
Analyzer module temperature not
in permissible range
258
Cause
Safety function
Ambient temperature too high
One or more heater control loops
faulty
Pressure sensor faulty
Reference gas open, sample gas
input/output closed
Reference gas pressure - sample
gas pressure < 2 000 hPa
Sample gas path closed
Reference gas path leaking
Zero gas changed, setpoint faulty
Incorrect reference gas
Drift probe
Calibration successful
Span gas changed, setpoint faulty
Incorrect reference gas
Drift probe
Calibration successful
Frequency change executed
Difference of O
concentration be-
2
tween zero gas and reference gas
too large.
Gas line dirty, gas cooler faulty
Gas lines leaking
Incorrect zero gas or setpoint
faulty
Sample gas concentration too
high
Sample gas pressure too
high/outside pressure correction
range
Device is operated outside the per-
mitted ambient temperature range
Remedy
Check ambient temperature
Check the other messages includ-
ing the logbook and act accord-
ingly
Set sample gas pressure to per-
mitted range.
If sample gas input/output is
closed, the reference gas must be
turned off.
Replace analyzer module
Check pressure of reference gas
source
Check flow of reference gas, see
operating instructions.
Check setpoint
Check reference gas
Replace analyzer module
Execute zero point calibration
Check setpoint
Check reference gas
Replace measuring head
Execute span calibration
Normalization required; contact Ser-
vice if necessary
Check zero gas, reference gas
pressure
Carry out zero and full-scale val-
ue calibrations
Check gas preparation, e.g.,
check filter
Check gas lines for leaks
Check setpoint
Check sample gas concentration
Check sample gas pressure
Execute span gas calibration
Check ambient temperature
Operating with the Local User Interface