5.4.1
Recommended test setup
5.4.1.1
Recommended test setup for analyzer modules in all versions
Arrangement of the gas connections in analyzer modules in all versions
The figure below is a schematic view of the gas connections:
1
2
3
4
Figure 5-1
SIPROCESS GA700 rack and wall mounted devices
Compact Operating Instructions, 06/2023, A5E35134047-AB
Sample gas inlet
Sample gas outlet
ULTRAMAT 7: Reference gas outlet
OXYMAT 7 (high-pressure version) / CALOMAT 7: Not assigned, dummy plug
OXYMAT 7 (low-pressure version): Bypass outlet
OXYMAT 7 / ULTRAMAT 7: Reference gas inlet
CALOMAT 7: Not assigned, dummy plug
Arrangement of the gas connections
Commissioning
5.4 Checking gas paths for leaks
69