Calibration Requirements; Overview; Permissibility Of Calibrations - Siemens SIPROCESS GA700 Operating Manual

Continuous gas analysis, operating with the local user interface
Hide thumbs Also See for SIPROCESS GA700:
Table of Contents

Advertisement

Functions
6.7 [2.08] Settings > Calibrations
6.7.2

Calibration requirements

6.7.2.1

Overview

The permissibility of a calibration start depends on which message(s) are currently active in
the device.
● An overview of the messages whose presence permits the start of a calibration is
available in section → Permissibility of calibrations (Page 118).
● The execution of a calibration is generally not permitted in the following cases:
– The device is in "Warming-up" mode.
– Active messages exist that are not covered by specific exceptions or will be canceled:
The device checks if execution of a calibration is permitted. If the calibration is not
permitted, a corresponding message is displayed and the process is canceled.
An overview of the messages whose presence makes the start of a calibration
impermissible is available in section → Non-permissibility of calibrations (Page 119).
6.7.2.2

Permissibility of calibrations

Overview
The tables below provide an overview of cases in which the execution of a calibration is
permitted or may be necessary to remove a message. Information on additional messages
and the corresponding remedial measures is available at:
● → OXYMAT 7 message list (Page 257)
● → ULTRAMAT 7 message list (Page 264)
● → CALOMAT 7 message list (Page 269)
Permitted calibration
Table 6- 4
Message
No.
-
1)
-
1)
Diverse, comparable message list.
1)
118
Permissibility of calibrations as part of the interference gas measurement
Text
An active message affects another component that is not calibrated.
Active messages created by the device are available.
Operating with the Local User Interface
Operating Manual, 06/2017, A5E31930478-05