Alarm, error, and system messages
7.4 CALOMAT 7 message list
Message
No.
Text
146
Configuration faulty
148
Saved user data faulty
149
Factory data faulty
150
Parameter value invalid
151
Configuration restored
following fault
153
Factory synchronization
of the electronics faulty
154
Saved sensor module
data faulty
226
Gas warning preliminary
alarm
227
Gas warning main alarm
238
Cross-interference meas-
ured value uncertain or
faulty
239
External pressure value
faulty
272
Cause
Content of parameter memory faulty
and cannot be restored
Content of user data in parameter
memory faulty and cannot be restored
Content of factory data in parameter
memory faulty and cannot be restored
Change of input limits of a parameter
due to change of another parameter
The previously set value is now out-
side the changed input limits
Content of parameter memory was
faulty but has been recovered
Synchronization values of analyzer
module/option module electronics
faulty and cannot be restored
Content of sensor module data in
parameter memory faulty and cannot
be restored
Gas warning (preliminary alarm)
Gas warning alarm (main alarm)
External or internal interference gas
measured value is faulty
Alternatively: the external interference
gas measured value is not transferred
External pressure measured value is
not being transferred or is faulty
Remedy
The active device configuration is
faulty
Load the user or factory data in the
work data.
Save the work data in the user data
again
Alternatively: Transfer a valid parame-
ter set to the device
The saved factory data is faulty
Contact Service
Due to a change in a parameter, the
value of another parameter is outside
its valid limits
Correct the value of the associated
parameter
An error was detected when loading
the device configuration
However, a valid configuration could
be recovered without data loss
The synchronization data saved at the
factory for the module electronics is
corrupt
Replace the module
The saved sensor data is faulty
Contact Service
Check the sources for the measured
values of the set interference gases.
Check the external pressure meas-
urement
Operating with the Local User Interface
Operating Manual, 06/2017, A5E31930478-05