Alarm, error, and system messages
7.4 CALOMAT 7 message list
Message
No.
Text
146
Configuration faulty
148
Saved user data faulty
149
Factory data faulty
150
Parameter value invalid
151
Configuration restored
following fault
153
Factory synchronization
of the electronics faulty
154
Saved sensor module
data faulty
226
Gas warning preliminary
alarm
227
Gas warning main alarm
238
Cross-interference meas-
ured value uncertain or
faulty
239
External pressure value
faulty
272
Cause
Content of parameter memory faulty
•
and cannot be restored
Content of user data in parameter
•
memory faulty and cannot be restored
Content of factory data in parameter
•
memory faulty and cannot be restored
Change of input limits of a parameter
•
due to change of another parameter
The previously set value is now out-
•
side the changed input limits
Content of parameter memory was
•
faulty but has been recovered
Synchronization values of analyzer
•
module/option module electronics
faulty and cannot be restored
Content of sensor module data in
•
parameter memory faulty and cannot
be restored
Gas warning (preliminary alarm)
•
Gas warning alarm (main alarm)
•
External or internal interference gas
•
measured value is faulty
Alternatively: the external interference
•
gas measured value is not transferred
External pressure measured value is
•
not being transferred or is faulty
Remedy
The active device configuration is
•
faulty
Load the user or factory data in the
•
work data.
Save the work data in the user data
•
again
Alternatively: Transfer a valid parame-
•
ter set to the device
The saved factory data is faulty
•
Contact Service
•
Due to a change in a parameter, the
•
value of another parameter is outside
its valid limits
Correct the value of the associated
•
parameter
An error was detected when loading
•
the device configuration
However, a valid configuration could
•
be recovered without data loss
The synchronization data saved at the
•
factory for the module electronics is
corrupt
Replace the module
•
The saved sensor data is faulty
•
Contact Service
•
Check the sources for the measured
•
values of the set interference gases.
Check the external pressure meas-
•
urement