Procedure
1. Main menu > "3. Maintenance & diagnostics" > "11. Test" > "1. Test inputs/outputs" >
"1. Analog outputs"
Use this menu to access the assigned test parameters of all analog outputs. You can also
start the test function for all activated analog outputs.
Note
Equipment-dependent display
Only the analog outputs actually available on the device are listed in the menu. Example:
• Analyzer module 1 with a component exists: Only analog output AO11 is displayed
• Analyzer modules 1 and 2 exist each with one component: The analog outputs AO11
Note
Principle of operation
When you enable the test function, the analog outputs are released from their actual
function.
The navigation lines of the menus are structured as follows:
1)
2)
3)
4)
5)
2. Enter setpoint for testing the first analog output, e.g.:
Main menu > "3. Maintenance & diagnostics" > "11. Test" > "1. Test inputs/outputs" >
"1. Analog outputs" > "2. AO01: ..."
3. Specify unit for setpoint to be defined: "Setpoint input" parameter field.
– Authorize access with Standard PIN, if necessary.
– Depending on the measuring device, select the value "mA" or "Gas unit".
4. Specify the test setpoint: "Setpoint in mA" or "Setpoint in gas unit" parameter field.
– When you output the setpoint in the gas unit, the analog output is output depending on
– Enter the corresponding value.
Operating with the Local User Interface
Operating Manual, 06/2017, A5E31930478-05
and AO21 are displayed.
Structure of the navigation lines
1.
Test mode
2.
AO11: O2
2) 3) 4)
Activity indicator test mode (active/ inactive)
Designation of the analog output and the component
AM1: AO11 (component 1), AO12 (component 2), AO13 (not assigned)
AM2: AO21 (component 3), AO22 (component 4), AO23 (not assigned)
Actual value display
the assigned measuring range.
6.24 [3.11] Maintenance & Diagnostics > Test
Specify setpoint for testing ►
Functions
Active
1)
Edit ►
4 mA
5)
237