Communications; Qualified Personnel; Automatic Reset - Siemens SIPROCESS GA700 Operating Manual

Continuous gas analysis, operating with the local user interface
Hide thumbs Also See for SIPROCESS GA700:
Table of Contents

Advertisement

2.3.3

Communications

Note
Communication protocols
The devices described in this documentation are able to communicate using Ethernet and
MODBUS TCP. The protocols used are basically non-secure in the sense of IT security.
• Make sure that only secure connections are used for communication.
• Only use the device within appropriately secure networks.
Note
Illegal Ethernet connection
Illegal Ethernet connections to the analyzer device can bypass firewalls.
Make sure that the analyzer device is not connected to or can be connected to two separate
Ethernet networks.
2.3.4

Qualified personnel

Qualified personnel are individuals who are familiar with the installation, mounting,
commissioning, and operation of the product. These individuals have the following
qualifications:
● They are authorized, trained or instructed in operating and maintaining devices and
systems according to the safety regulations for electrical circuits, high pressures, and
aggressive as well as dangerous media.
● For explosion-proof devices: They are authorized, trained, or instructed in working on
electrical circuits for systems in hazardous areas.
● They are trained or instructed in maintenance and use of appropriate safety equipment
according to the safety regulations.
2.3.5

Automatic reset

All SIPROCESS GA700 devices feature internal monitoring. When this monitoring is
activated, the standard digital outputs switch after about 4 seconds for a period of approx. 60
seconds to "inactive", i.e. to a recognized safe operating mode. These times are valid for
devices manufactured as of 09/2017. You can find this information on the nameplate. It can
take up to 10 seconds for older devices to switch over, and up to 100 seconds to switch to
safe operating mode.
Operating with the Local User Interface
Operating Manual, 06/2017, A5E31930478-05
General information
2.3 Information on use
17

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents