Alarm, error, and system messages
7.2 OXYMAT 7 message list
Message
No.
Text
82
Reference voltage outside toler-
ance
83
Signal acquisition faulty
84
Test signal outside tolerance
85
System error
86
System error
87
AM parameter memory faulty
88
Analog output faulty
89
ADC faulty
90
Program memory faulty
91
Data storage faulty
92
Data storage faulty
93
Real-time clock faulty
128
SPI communication faulty
129
Communication with ADU faulty
130
CAN communication faulty
131 to
Communication error with
137
MODBUS TCP nodes 1 to 7
260
Cause
Electronic fault
Electronic fault
Electronic fault
Electronic fault
Electronic fault
Electronic fault
Electronic fault
Electronic fault
Electronic fault
Electronic fault
Electronic fault
Electronic fault
AM electronics faulty
AM electronics faulty
CAN bus connection faulty
•
AM electronics faulty
•
PU electronics faulty
•
Assigned IP address of
•
MODBUS TCP node 1 faulty.
Ethernet connection faulty (con-
•
nector, cable)
In case of gateway via router:
•
Assigned gateway address is
faulty
Remedy
Replace analyzer module
Replace analyzer module
Replace analyzer module
Restart the device
Restart the device
Replace analyzer module
Check circuit
•
Replace analyzer module
•
Replace analyzer module
Replace analyzer module
Replace analyzer module
Replace basic unit
Replace basic unit
Replace analyzer module
Replace analyzer module
Check plug-in connections of
•
CAN cable on the modules
If a specific analyzer module is
•
not recognized during startup ⇒
replace module.
If no analyzer module is recog-
•
nized during startup ⇒ replace
basic unit.
Check assigned IP address
•
Check Ethernet plug-in connec-
•
tions and cables
In case of gateway via router:
•
Check assigned default gateway
address