Siemens SIPROCESS GA700 Operating Manual page 263

Continuous gas analysis, operating with the local user interface
Hide thumbs Also See for SIPROCESS GA700:
Table of Contents

Advertisement

Message
No.
Text
237
AutoCal 2 canceled
238
Cross-interference measured
value uncertain or faulty
239
External pressure value faulty
Operating with the Local User Interface
Operating Manual, 06/2017, A5E31930478-05
Cause
AutoCal could not be executed due to
a fault.
External or internal cross-interference
measured value is faulty or the exter-
nal cross-interference measured
value is not transmitted.
External pressure value is not trans-
mitted or is faulty.
Alarm, error, and system messages
7.2 OXYMAT 7 message list
Remedy
Check and remedy pending errors
in the list of current messages.
Then: Execute AutoCal manually.
Check the sources for the measured
values of the parameterized interfer-
ence gases.
Check source of measured value
for assignable interference gases
Check interference gas supply
If transmission is via MODBUS
TCP, check if errors no. 131 to
137 are pending and follow their
instructions.
Check external pressure meas-
urement
If transmission is via MODBUS
TCP, check if errors no. 131 to
137 are pending and follow their
instructions.