[3.05.3] Electric Parameters - Siemens SIPROCESS GA700 Operating Manual

Continuous gas analysis, operating with the local user interface
Hide thumbs Also See for SIPROCESS GA700:
Table of Contents

Advertisement

Only ULTRAMAT 7: predictive self-diagnostics
Table 6- 58
Parameter
Source intensity
Intensity I
6.19.3

[3.05.3] Electric parameters

Overview
In this menu you access information/current values that you can use to evaluate the supply
status of the device. The menu shows the modules installed in the device for which electrical
parameters are available. This includes analyzer modules and any installed option modules.
The menu can include up to three navigation lines, depending on the features:
You only have read access.
Navigation lines in the menu [3.05.03]
Navigation number
AM1 → "1.", AM2 → "2.", OM2 → "4."; "3" currently not assigned.
1)
Depending on features: AM1, AM2, OM2
2)
Gas component or module type, for OM2, e.g., "AI/DI"
3)
Operating with the Local User Interface
Operating Manual, 06/2017, A5E31930478-05
ULTRAMAT 7: [3.05.2] Predictive self-diagnostics
High alarm limit
High warning limit
Current deviation Percentage deviation of the current intensity I
Low warning limit Maximum negative deviation of the source intensity (in %). A
Low alarm limit Maximum negative deviation of the source intensity (in %). A
TC
0
Module name
1)
6.19 [3.05] Maintenance & Diagnostics > Diagnostics values
Description
Maximum positive deviation of the source intensity (in %). A
maintenance alarm is signaled on a high limit violation.
Maximum positive deviation of the source intensity (in %). A
maintenance demanded is signaled on a high limit violation.
normalized intensity I
maintenance demanded is signaled on a low limit violation.
maintenance alarm is signaled on a low limit violation.
Source intensity I
after temperature compensation
0
Component/ module
2)
type
3)
0
.
0 norm
Display details ►
Functions
from the
219