Description
3.6 Combined operation
Note
Installation of analyzer modules
The analyzer module is only installed in the device at the factory when you select the options
"D00 to D99" when ordering an analyzer module.
If you do not specify the module assignment/installation location during the ordering process,
the analyzer module is delivered separately from the device (supplied separately).
If the analyzer module is supplied separately, you must install it yourself. You will find the
information on installation only in the Operating Instructions of the device: → References
(Page 281).
NOTICE
Installation of analyzer modules
Analyzer modules that are installed in a device version other than the one for which they
were intended can cause damage. Therefore:
• Prior to installation, check the device version for which the analyzer module is intended.
• Only install analyzer modules manufactured for operation in a rack-mounted device into
• Only install analyzer modules manufactured for operation in a wall-mounted device into
Additional information is also available from your Siemens partner → Technical support
(Page 279).
Note
Influence of building vibrations on the measured signal
If you are using OXYMAT 7 and/or ULTRAMAT 7 analyzer modules, building vibrations, in
particular in the frequency range from 6 to 15 Hz, can influence the measured signal.
• Select an installation location which is vibration-free.
• If necessary, change the measuring frequency settings:
– Make sure that the measuring frequencies do not correspond to the building vibration
– Also make sure that the measuring frequency is not the same as half the fundamental
32
You will find the relevant information in the article number on the module nameplate:
– 7MB3010-xxxxx-xxAx-Z → "A": Analyzer module for rack-mounted devices
– 7MB3010-xxxxx-xxBx-Z → "B": Analyzer module for wall-mounted devices
such a device.
such a device.
or to one of its harmonic multiples.
frequency of the building vibration.