Siemens SIPROCESS GA700 Operating Manual page 286

Continuous gas analysis, operating with the local user interface
Hide thumbs Also See for SIPROCESS GA700:
Table of Contents

Advertisement

Appendix
A.5 MODBUS TCP
Status byte
A status byte is additionally transmitted in synchronism with each measured value. This byte
provides information on the quality of the value in the form of a "Status code". The status
byte has the following structure:
Table A- 8
Coding of status byte/structure
Quality
Substatus
Bit 7
Bit 6
Bit 5
0
0
0
1
1
0
With limit violations within one band
1)
Coding of the status byte for the measured value output
Table A- 9
Measured value status "Good" (MODBUS TCP)
Quality byte
7 6 5 4 3 2 1 0
1 0 0 0 1 1 0 1 141
1 0 0 0 1 1 1 0 142
1 0 0 0 1 1 1 1 143
1 0 0 0 1 0 0 1 137
1 0 0 0 1 0 1 0 138
1 0 0 0 1 0 1 1 139
1 0 1 1 1 1 0 0 188
1 0 1 0 1 0 0 0 168
1 0 1 0 0 1 0 0 164
1 0 0 0 0 0 0 0 128
286
Bit 4
Bit 3
Dec
Hex
Cause
0x8d
The measured value is valid, but is below an alarm limit.
0x8e
The measured value is valid, but is above an alarm limit.
0x8f
The measured value is valid, but is within an invalid alarm
limit band.
0x89
The measured value is valid, but is below a warning limit.
0x8a
The measured value is valid, but is above a warning limit.
0x8b
The measured value is valid, but is within a warning limit
band.
0xbc
The measured value is valid, but the function check is ac-
tive.
For example, measuring mode is active and a user is logged
on.
0xa8
The measured value is valid, but a maintenance demanded
is signaled.
0xa4
The measured value is valid, but a maintenance demanded
is signaled.
0x80
The measurement is valid (normal measuring mode).
Limits
Bit 2
Bit 1
Bit 0
0
0
0
1
1
0
1
1
Operating Manual, 06/2017, A5E31930478-05
Structure
bad
uncertain
good
ok
low limited
high limited
high and low limited
Effect/remedy
-
-
-
-
-
-
-
-
-
-
Operating with the Local User Interface
Status
1)