Siemens SIPROCESS GA700 Operating Manual page 125

Continuous gas analysis, operating with the local user interface
Hide thumbs Also See for SIPROCESS GA700:
Table of Contents

Advertisement

Depending on the hardware of the device, the sample gas purging time can be set device-
specific or module-specific:
Table 6- 12
Device-specific
If only one analyzer module is installed or if a
common gas path exists for two analyzer mod-
ules (series connection).
Check the setup of the gas path as well as the parameter assignment of the process tags
before you set the sample gas purging time: [2.12] Settings > Gas path/process tag label
(Page 183).
Procedure
1. Main menu > "2. Settings" > "08. Calibrations" > "1. Setpoints, tolerances, purge
time" > "4. Sample gas purging time"
2. Specify duration: "Sample gas purging time" parameter box.
– Authorize access with Expert PIN, if necessary.
– Increase/reduce preset value.
– During the selected time, the "Function check" remains set. After the purge time has
Operating with the Local User Interface
Operating Manual, 06/2017, A5E31930478-05
Effect of the sample gas purging time
elapsed, the device automatically changes to the "Measurement" operating state.
6.7 [2.08] Settings > Calibrations
Module-specific
If two analyzer modules with separate gas paths
are installed, set the parameters separately for
each analyzer module.
Functions
125