Plug & Measure Permitted Activities; Plug & Measure - Siemens SIPROCESS GA700 Operating Manual

Continuous gas analysis, operating with the local user interface
Hide thumbs Also See for SIPROCESS GA700:
Table of Contents

Advertisement

Description
3.6 Combined operation
Retrofitting example "Suction operation, serial" operating mode for OXYMAT 7/ULTRAMAT 7
When you are using the ULTRAMAT 7 and the OXYMAT 7 analyzer module in serial suction
operation, you must convert the sample gas restrictor of the OXYMAT 7 analyzer module:
Sample gas inlet → Sample gas outlet.
MGE, VGE
MGA/VGA
Figure 3-12
3.6.4
Plug & measure permitted activities
3.6.4.1
Plug & measure
Term
"Plug & Measure" is the ability to recognize any analyzer modules and/or option modules
added at a later time and to integrate them. This ability makes it easier to replace modules
for maintenance purposes as well as installing devices at a later time, for example, option
modules.
When you replace or retrofit a module, the processing module recognizes the change in the
device configuration during a restart. The type of module as well as its parameter
assignment are recognized. You can either apply the parameter assignments of the
replacement/retrofit module or use those of the previous module.
44
Sample gas inlet (MGE), Reference gas inlet (VGE)
Sample gas outlet (MGA)/reference gas outlet
Remove sample gas restrictor from the sample gas inlet of the additional OXYMAT 7
analyzer module
Install the clamping screw into the sample gas inlet of the additional OXYMAT 7 ana-
lyzer module
Install the sample gas restrictor into the sample gas outlet of the additional OXYMAT 7
analyzer module
Install additional OXYMAT 7 analyzer module
Target configuration: ULTRAMAT 7 and OXYMAT 7 in serial suction operation
Retrofitting of the sample gas restrictors: Example "Pressure operation, serial"
Operating with the Local User Interface
Operating Manual, 06/2017, A5E31930478-05