1. "1 Quick Start" > "4. Measuring ranges" > "n component" main menu
The menu includes four navigation lines for setting the measuring ranges and one
navigation line for setting autoranging.
2. Call autoranging: Navigation line "Autoranging".
The parameter display of autoranging has a parameter field and a display box.
3. Specify measuring range: "Measuring range selection" parameter field.
– Authorize access with Standard PIN, if necessary.
– Select one of the offered measuring ranges or autoranging:
8.7

[1.5] Calibration

Procedure
Note
Component-specific setup
Measuring ranges are component-specific. Repeat the procedure described below for all
components, if necessary.
1. "1 Quick Start" > "4. Measuring ranges" > "n component" main menu
The menu includes a navigation line for setting of autoranging.
One navigation line is reserved for each of the four measuring ranges that can be set:
1)
2. Open selected measuring range: Press <ENTER>.
The parameter display of the measuring range is opened.
Quick Start
Compact Operating Instructions, 06/2017, A5E31805656-05
If the message "Active" is displayed in the "Autoranging" line, autoranging is activated.
If "Not possible" is displayed, correct the start and end values of the individual
measuring ranges: Setting the measuring ranges (Page 96).
Structure of the navigation line for each measuring range
Measuring range
Measuring range start - measuring range end
1 to 4
On/Off: signals if the measuring range is currently active and in use.
Commissioning: Sequence
8.7 [1.5] Calibration
Activation status
1)
Edit