Performance Specifications

Performance Specifications G1314B
Table 3
Type
Detection type
Light source
Wavelength range
Short term noise (ASTM)
Drift
Linearity
Wavelength accuracy
Maximum data rate
Band width
Flow cells
Agilent 1260 Infinity VWD User Manual
Performance Specifications G1314B
Specification
Double-beam photometer
Deuterium lamp
190 – 600 nm
<± 0.5·10
3·10
> 2 AU (5 %) upper limit
± 1 nm
13 Hz
6.5 nm typical
Standard: 14 µL volume,
10 mm cell path length and
40 bar (580 psi) pressure
maximum
High pressure: 14 µL volume,
10 mm cell path length and
400 bar (5800 psi) pressure
maximum
Micro: 1 µL volume, 5 mm cell
path length and 40 bar
(580 psi) pressure maximum
Semi-micro: 5 µL volume,
6 mm cell path length and
40 bar (580 psi) pressure
maximum
Site Requirements and Specifications
Performance Specifications
-5
AU at 254 nm
-4
AU/h at 254 nm
Comments
See
"Specification
Conditions"
on page 29
See
"Specification
Conditions"
on page 29.
See
"Specification
Conditions"
on page 29.
Self-calibration with
deuterium lines, verification
with holmium oxide filter
Can be repaired on component
level
2
25