To test the time interval accuracy (logic analyzer)

Testing the time interval accuracy does not check a specification, but does check the
following:
125 MHz oscillator
This test verifies that the 125 MHz timing acquisition synchronizing oscillator is
operating within limits.
Equipment Required
Equipment
Critical Specifications
Pulse Generator
100 MHz 3.5 ns pulse width, < 600 ps rise time
Accuracy ≤ (5)(10
Function Generator
SMA Cable
Adapter
BNC(m)-SMA(f)
BNC Test Connector,
6x2

Set up the equipment

Turn on the equipment required and the logic analyzer. Let them warm up for
1 1
30 minutes if you have not already done so.
Set up the pulse generator according to the following table.
2 2
Pulse Generator Setup
Channel 1
Period
Delay: 0 ps
5 ns
Width: 2.5 us
High: − 0.9 V
Low: − 1.7 V
COMP: Disabled
(LED off)
− 6
) × frequency
Mode
EXT TRIG
TRIG
Slope: Positive
THRE: 1.0 V
Recommended
Model/Part
HP 8131A Option 020
HP 3325B Option 002
HP 8120-4948
HP 1250-2015
3–59