NOTE
Figure 2-6
Make a Transmission Measurement
1. Connect the equipment as shown in
cable such as the one that was supplied with your analyzer.
The quality of the cable will affect these measurements; make sure you
use a cable with the characteristics described in
page
2-10.
Equipment Setup for Performing the Operator's Check
2. Press
PRESET
Enter
.
3. Press
POWER
4. Press
Default Response
CAL
5. Verify that the data trace falls within ±0.5 dB of 0 dB. See
for a typical HP 8714ET result. The HP 8712ET should look similar,
but end at 1300 MHz.
ET User's Guide
Figure
User Preset
Factory Preset
or
dBm
.
0
.
Getting Started
Performing the Operator's Check
2-6. Use a known good
"Equipment List" on
SCALE
Figure 2-7
.1