Table 9-5
Test Port Output (continued)
Output Power (continued)
Maximum Power Setting:
HP 8712ET:
50 Ω:
With Attenuator Option 1E1
Without Attenuator Option 1E1
75 Ω:
With Attenuator Option 1E1
Without Attenuator Option 1E1
HP 8712ES:
50 Ω
75 Ω
HP 8714ET:
50 Ω:
With Attenuator Option 1E1
Without Attenuator Option 1E1
75 Ω:
With Attenuator Option 1E1
Without Attenuator Option 1E1
HP 8714ES:
50 Ω
75 Ω
a. Maximum power setting is the highest power setting allowed that maintains the network
analyzer's specified level accuracy.
b. Option 1E1 adds a 60 dB step attenuator to the HP 8712ET/8714ET.
HP 8712ET/ES and HP 8714ET/ES Test Port Output
Description
a
b
b
b
b
ET User's Guide
Specification
<1 GHz
>1 GHz
15 dBm
12 dBm
16 dBm
13 dBm
12 dBm
9 dBm
13 dBm
10 dBm
13 dBm
10 dBm
10 dBm
7 dBm
10 dBm
9 dBm
11 dBm
10 dBm
7 dBm
6 dBm at
<2 GHz
8 dBm
7 dBm at
<2 GHz
9 dBm
7 dBm
6 dBm
4 dBm at
<2 GHz
Specifications
Supplemental
Information
6 dBm at >2 GHz,
char.
7 dBm at >2 GHz,
char.
4 dBm at >2 GHz,
char.