Figure 3-10
Short Connection to the Adapter
7. Select the delay format for your measurement by pressing
8. Press
9. Press
10. Press
11. Divide the mean value by 2. Record this value for use later in the
12. Select a type-N(f) calibration kit for the REFLECTION port. This is
ET User's Guide
FORMAT
Delay
to automatically select marker 1 as the active
MARKER
marker. Position marker 1 to the left edge of the measurement trace.
to select marker 2 as the active marker. Position marker 2
2:
to the right edge of the measurement trace.
Marker Functions Marker Math Statistics
mean value, located in the upper right corner of the display.
procedure.
necessary because the type-N(m) short (connected to the adapter on
port 2) is a component in a type-N(f) calibration kit. Press
More Cal Cal Kit
keys to highlight the REFLECTION port in the display's table.
Press
Type-N(f)
.
Measuring Transmission Response
.
, and use the front panel knob or the
Making Measurements
. Find the
CAL
3-27