Calibration Choices

To choose the appropriate calibration method suited to your type of
measurement, follow these steps:
1. Set up the analyzer for your measurement:
• Select
MEAS 1
• Enter operating parameters other than the default.
2. When performing a one-port reflection calibration or an enhanced
response calibration, you must select a calibration kit to match the
port at your calibration reference plane. Refer to
Reference Plane" on page 6-5
3. Press the
CAL
measurement you selected in step 1.
See
Table 6-1
and
Figure 6-4
calibrations available for each type of measurement. Later sections in
this chapter discuss these different types of calibrations, organized by
measurement type. There is also a discussion of a type of calibration
called normalization.
ET User's Guide
Calibrating for Increased Measurement Accuracy
or
.
MEAS 2
and
"Calibration Kits" on page
key to bring up the calibration choices for the type of
on the following pages for the types of
Calibration Choices
"The Calibration
6-16.
6-7