Table 9-6
Test Port Input (continued)
Noise Floor
HP 8712ET/8714ET:
50 Ω:
Narrowband
Broadband (Internal)
75 Ω:
Narrowband
Broadband (Internal)
HP 8712ES/8714ES:
50 Ω:
Narrowband
Broadband (Internal)
75 Ω:
Narrowband
Broadband (Internal)
a. Noise floor is defined as the RMS value of the trace (in linear format) for a trans-
mission measurement in CW Frequency mode, with the RF connectors termi-
nated in loads, output power set to 0 dBm, and no averaging. The noise floor
specifications and typicals for narrowband detection measurements assume that
an isolation calibration has been performed using an average factor of 16. For
the HP 8712ES/8714ES, external broadband detectors will provide a lower noise
floor than the internal broadband detectors.
HP 8712ET/ES and HP 8714ET/ES Test Port Input
Description
(in dBm)
Fine
(15 Hz)
a
−105
−50
−104
−47
−96
−38
−95
−35
ET User's Guide
Spec.
(in dBm)
System Bandwidths:
Fine
Med Wide
(15 Hz)
(4000 Hz)
−111
−86
−55
−48
−109
−84
−52
−45
−105
−80
−43
−36
−104
−80
−40
−33
Typical
Wide
(6500 Hz)
−48
−32
−46
−30
−47
−23
−47
−20
Specifications
9-21