Configure process measurement
Procedure
Set Volume Flow Type to Liquid.
4.2.2
Configure Volume Flow Measurement Unit for liquid
applications
Display
ProLink II
ProLink III
Field Communicator Configure > Manual Setup > Measurements > Flow > Volume Flow Unit
Overview
Volume Flow Measurement Unit specifies the unit of measurement that will be displayed for the
volume flow rate. The unit used for the volume total and volume inventory is based on this
unit.
Prerequisites
Before you configure Volume Flow Measurement Unit, be sure that Volume Flow Type is set to
Liquid.
Procedure
Set Volume Flow Measurement Unit to the unit you want to use.
The default setting for Volume Flow Measurement Unit is l/sec (liters per second).
Tip
If the measurement unit you want to use is not available, you can define a special measurement unit.
Options for Volume Flow Measurement Unit for liquid
applications
The transmitter provides a standard set of measurement units for Volume Flow Measurement
Unit, plus one user-defined measurement unit. Different communications tools may use
different labels for the units.
Table 4-3:
Options for Volume Flow Measurement Unit for liquid applications
Unit description
Cubic feet per second
Cubic feet per minute
30
OFF-LINE MAINT > OFF-LINE CONFG > UNITS > VOL
ProLink > Configuration > Flow > Vol Flow Units
Device Tools > Configuration > Process Measurement > Flow
Display
ProLink II
CUFT/S
ft3/sec
CUF/MN
ft3/min
Micro Motion
Label
ProLink III
ft3/sec
ft3/min
®
Model 2700 Transmitters with Analog Outputs
Field Communica-
tor
Cuft/s
Cuft/min