View And Acknowledge Alarms Using Prolink Ii - Emerson Micro Motion 2700 Configuration And Use Manual

Transmitters with analog outputs
Hide thumbs Also See for Micro Motion 2700:
Table of Contents

Advertisement

Postrequisites
9.4.2

View and acknowledge alarms using ProLink II

You can view a list containing all alarms that are active, or inactive but unacknowledged.
From this list, you can acknowledge individual alarms.
1.
2.
3.
Postrequisites
Configuration and Use Manual
To clear the following alarms, you must correct the problem, acknowledge the
alarm, then power-cycle the transmitter: A001, A002, A010, A011, A012, A013,
A018, A019, A022, A023, A024, A025, A028, A029, A031.
For all other alarms:
-
If the alarm is inactive when it is acknowledged, it will be removed from the list.
-
If the alarm is active when it is acknowledged, it will be removed from the list
when the alarm condition clears.
Choose ProLink > Alarm Log.
Choose the High Priority or Low Priority panel.
Note
The grouping of alarms into these two categories is hard-coded and is not affected by Status
Alarm Severity.
All active or unacknowledged alarms are listed:
Red indicator: Alarm is currently active.
Green indicator: Alarm is inactive but unacknowledged.
Note
Only Fault and Informational alarms are listed. The transmitter automatically filters out alarms
with Status Alarm Severity set to Ignore.
To acknowledge an alarm, check the Ack checkbox.
To clear the following alarms, you must correct the problem, acknowledge the
alarm, then power-cycle the transmitter: A001, A002, A010, A011, A012, A013,
A018, A019, A022, A023, A024, A025, A028, A029, A031.
For all other alarms:
-
If the alarm is inactive when it is acknowledged, it will be removed from the list.
-
If the alarm is active when it is acknowledged, it will be removed from the list
when the alarm condition clears.
Transmitter operation
155