Set Up The Weights & Measures Application Using Prolink Iii - Emerson Micro Motion 2700 Configuration And Use Manual

Transmitters with analog outputs
Hide thumbs Also See for Micro Motion 2700:
Table of Contents

Advertisement

8.3
Set up the Weights & Measures application
using ProLink III
When your transmitter is set up to comply with OIML or NTEP requirements, and the
transmitter is "secured," selected measurement data from the transmitter is approved for
Weights & Measures applications.
Prerequisites
Important
Review the Weights & Measures requirements for your location and ensure that you are complying
with all local requirements as well as the OIML or NTEP requirements.
If necessary for your installation, arrange for a site visit by a certified Weights & Measures
inspector, and ensure that the inspector is present for the appropriate portions of this
procedure.
Ensure that the transmitter is ready to be secured, i.e., it is configured as desired and you
have performed all appropriate tests and adjustments. After the transmitter is secured,
you cannot make any changes to configuration, many maintenance actions are disallowed,
and some operator actions are disallowed.
Procedure
1.
2.
3.
Configuration and Use Manual
Make a service port connection from ProLink III to your transmitter.
Choose Device Tools > Configuration > Weights & Measures and set Regulatory Agency to the
apropriate regulatory agency for your application.
Option
NTEP
OIML
Choose Device Tools > Configuration > Totalizer Control Methods and set totalizer control
methods as desired.
Parameter
Reset Totalizers from Display
Reset Totalizers via Remote
Communications
Set up the Weights & Measures application
Description
Regulatory agency for the U.S.A. and Canada
Regulatory agency for all other world areas
Option
Enabled: You can reset totalizers from the display,
whether or not the transmitter is secured.
Disabled: You can reset totalizers from the display only
when the transmitter is unsecured.
Enabled: You can reset totalizers using digital communi-
cations, whether or not the transmitter is secured.
Disabled: You can reset totalizers using digital communi-
cations only when the transmitter is unsecured.
143

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents