Configure process measurement
7.
Concentration process variables are now available on the transmitter. You can view and
report them in the same way that you view and report other process variables.
4.8.4
Standard matrices for the concentration measurement
application
The standard concentration matrices available from Micro Motion are applicable for a
variety of process fluids.
See
Table 4-14
Micro Motion, along with the density and temperature measurement units used in
calculation, and the unit used to report concentration data.
66
Option
Setup
Polling for tempera-
a. Ensure that the primary mA output has been wired to support
ture
b. Choose Online > Configure > Manual Setup > Measurements > External
c. Enable External Temperature.
d. Choose Online > Configure > Manual Setup > Measurements > External
e. Set Poll Control to Poll As Primary Host or Poll as Secondary Host.
f. Choose an unused polling slot.
g. Set External Tag to the HART tag of the external temperature de-
h. Set Polled Variable to Temperature.
Tip
a. Choose Online > Configure > Manual Setup > Measurements > External
A value written by
digital communica-
tions
b. Enable External Temperature.
c. Perform the necessary host programming and communications
Note
If the Weights & Measures application is implemented and the trans-
mitter is secured, digital communications cannot be used to write
temperature data to the transmitter.
Online > Configure > Manual Setup > Measurements > Conc Measurement (CM) > CM
Configuration and set Active Matrix to the matrix to be used for measurement.
for a list of the standard concentration matrices available from
HART polling.
Pressure/Temperature > Temperature.
Pressure/Temperature > External Polling.
vice.
Poll as Primary: No other HART masters will be on the network.
Poll as Secondary: Other HART masters will be on the network.
The Field Communicator is not a HART master.
Pressure/Temperature > Temperature.
setup to write temperature data to the transmitter at appropri-
ate intervals.
®
Micro Motion
Model 2700 Transmitters with Analog Outputs