Installing a Thermostattable Cell Holder

1 Loosen the hold- down screws and remove the current sample holder
2 Orient the thermostattable cell holder the same way the standard cell
3 Tighten the hold- down screws.
4 Connect tubing (silicon, 3/16 inch i.d.) to the inlet and outlet of the
5 Insert the sample cell and lower the lever down to lock the cell in
6 Begin pumping thermostatted water through the manifold. Wait for the
When the thermostattable cell holder is not in use, it is best to disconnect
the tubing and drain the manifold.
Agilent Cary 8454 UV-Visible Spectroscopy System Installation Guide
from the sample compartment.
holder was oriented and lower it onto the sample compartment
mounting guides.
manifold. It does not matter which side is used as the inlet and which
side is used as the outlet. If necessary, route the tubing under the
tubing restraint tab to keep it from interfering with the light beam.
Make sure the tubing is securely attached; it expands when heated and
if it is not secure, a leak will result.
place.
sample cell and sample to reach thermal equilibrium before you make
any measurements. The time required to reach thermal equilibrium will
vary depending on the water temperature and the sample but generally
should take about five minutes.
Installing Accessories
Thermostattable Cell Holder
4