Agilent Technologies Cary 8454 Installation Manual page 26

Uv-visible spectroscopy system
Hide thumbs Also See for Cary 8454:
Table of Contents

Advertisement

2
Installation and Start Up
Installing your UV-Visible Spectrophotometer
• The multi- cell connector allows you to connect the cable which comes
• The GPIO (general purpose input/output) connector allow you to
• The remote connector may be used in combination with other analytical
• The RS 232 connector may be used to control the spectrophotometer
• The GPIB connector is used to connect the spectrophotometer with a
• The MIO slot for the Lan interface.
• The accessory board slot is reserved for future use.
• The power input socket accepts a line voltage of 100–240 V AC with a
On the right side of the instrument there is a door for exchanging the
lamps. Behind this plastic door there is another sheet metal door. Two
independent safety light switches are implemented. They automatically
turn off the lamps when the sheet metal door is opened.
26
from the multicell transport.
connect your sipper and autosampler or other accessories.
instruments from Agilent Technologies if you want to use features such
as common shut down, prepare etc.
from a computer through RS 232 connection, using appropriate software
(for future use). This connector needs to be activated by the
configuration switch module next to the GPIB connector. The software
needs the appropriate drivers to support this communication. See your
software documentation for further information.
computer. The 8- bit configuration switch module next to the GPIB
connector determines the GPIB address of your spectrophotometer. The
switches are preset to a default address recognized by the operating
software from Agilent Technologies.
line frequency of 50 or 60 Hz. Maximum power consumption is 220 VA.
There is no voltage selector on your spectrophotometer because the
power supply has wide ranging capability. There are no externally
accessible fuses, because automatic electronic fuses are implemented in
the power supply. The security lever at the power input socket prevents
that the spectrophotometer cover is taken off when line power is still
connected.
Agilent Cary 8454 UV-Visible Spectroscopy System Installation Guide