Site Requirements For Your Spectrophotometer; Power Considerations - Agilent Technologies Cary 8454 Installation Manual

Uv-visible spectroscopy system
Hide thumbs Also See for Cary 8454:
Table of Contents

Advertisement

Site Requirements for Your Spectrophotometer

Power considerations

Your spectrophotometer can operate on a line voltage of 100–240 V AC
with a line frequency of 50 or 60 Hz. Maximum power consumption is
220 VA.
The instrument power supply accepts any line voltage supplied in the
above mentioned range. Consequently there is no voltage selector in the
rear of the spectrophotometer. There are no externally accessible fuses,
because automatic electronic fuses are implemented in the power supply.
To disconnect the instrument from line power, disconnect the power cord from the
WA R N I N G
supply. The power supply still uses some power, even if the power switch on the
front panel is turned off.
The computer and InkJet or LaserJet printers accept a nominal line
voltage of 110 or 220 V AC with a line frequency of 50 or 60 Hz. For
more information, see the documentation that comes with your computer
and/or printer.
Connecting your instrument to a line voltage that is higher than specified causes a
WA R N I N G
shock hazard, and damage may occur to your instrument.
Agilent Cary 8454 UV-Visible Spectroscopy System Installation Guide