Site Requirements for Your Spectrophotometer
Power considerations
Your spectrophotometer can operate on a line voltage of 100–240 V AC
with a line frequency of 50 or 60 Hz. Maximum power consumption is
220 VA.
The instrument power supply accepts any line voltage supplied in the
above mentioned range. Consequently there is no voltage selector in the
rear of the spectrophotometer. There are no externally accessible fuses,
because automatic electronic fuses are implemented in the power supply.
To disconnect the instrument from line power, disconnect the power cord from the
WA R N I N G
supply. The power supply still uses some power, even if the power switch on the
front panel is turned off.
The computer and InkJet or LaserJet printers accept a nominal line
voltage of 110 or 220 V AC with a line frequency of 50 or 60 Hz. For
more information, see the documentation that comes with your computer
and/or printer.
Connecting your instrument to a line voltage that is higher than specified causes a
WA R N I N G
shock hazard, and damage may occur to your instrument.
Agilent Cary 8454 UV-Visible Spectroscopy System Installation Guide
Installation and Start Up
Site Requirements for Your Spectrophotometer
2
19