6
GPIB Communication, Connections and Configuration
Configuring the Agilent 82350 GPIB Interface
7 In the Agilent Connection Expert main window select GPIB0
Figure 37
8 Click Change Properties.... In the Agilent 82350 PCI GPIB Interface -
Figure 38
The appearance of the dialog slightly differs depending on the type of GPIB board installed.
N O T E
92
Figure
37).
Agilent 82350 default settings.
GPIB0 dialog window select or enter the following properties:
VISA interface ID: to GPIB0,
GPIB address: to 30,
check System controller (recommended),
SICL Interface ID: to hp82341 (note that this is case sensitive),
Logical unit: to 7 or 9 and
uncheck Auto- discover instruments connected to this interface
Figure
38).
Agilent 82350 interface configuration.
Agilent Cary 8454 UV-Visible Spectroscopy System Installation Guide
(see
(see