Agilent Technologies Cary 8454 Installation Manual page 124

Uv-visible spectroscopy system
Hide thumbs Also See for Cary 8454:
Table of Contents

Advertisement

7
LAN Communication, Installation, Connection and Configuration
Using the Agilent BootP Service
Using the Launch Manager
1 To add the instrument to the network go to Start > Programs >
2 Uncheck Do you want to log BootP requests?
3 Click Launch Manager. The BootP Manager screen appears.
Figure 60
4 Click Add.... The Add BootP Entry screen appears.
5 Make the following entries for your instrument:
6 Click OK.
7 Exit Launch Manager and power cycle both the PC and the
8 Ping the IP address from a command prompt on the PC to verify.
124
Agilent BootP Service>Edit BootP Settings. The BootP Settings screen
appears.
Configured interfaces view of the Agilent BootP Service
MAC Address
Host Name
IP Address
Comment, if desired
Subnet Mask
Gateway Address
spectrophotometer.
Agilent Cary 8454 UV-Visible Spectroscopy System Installation Guide