Installing The Cell Stirring Module - Agilent Technologies Cary 8454 Installation Manual

Uv-visible spectroscopy system
Hide thumbs Also See for Cary 8454:
Table of Contents

Advertisement

Installing the Cell Stirring Module

1 Remove the thermostattable cell holder from the sample compartment.
2 Remove the three screws that hold the thermostattable cell holder to its
3 Insert the cell stirring module into the cavity in the base of the
Figure 15
4 Tighten the four screws (flathead, M3×6 mm) that hold the cell stirring
5 Attach the thermostattable cell holder to its base.
6 Mount the thermostattable cell holder back in the sample compartment.
Agilent Cary 8454 UV-Visible Spectroscopy System Installation Guide
base. Remove the cell holder and turn it upside- down.
thermostattable cell holder, see
Installing the cell stirring module
module in place.
Installing Accessories
Cell Stirring Module
Figure
15.
4
55