Changing The Gpib Address Of The Spectrophotometer - Agilent Technologies Cary 8454 Installation Manual

Uv-visible spectroscopy system
Hide thumbs Also See for Cary 8454:
Table of Contents

Advertisement

Changing the GPIB Address of the Spectrophotometer

This section describes how to set/change the GPIB address of the
spectrophotometer.
GPIB addresses are only applied in configurations using a GPIB connection to the
N O T E
spectrophotometer.
In a LAN connection the GPIB address is not used.
The connector for the interface is located on the rear panel of the
spectrophotometer. Next to the connector is an address and control switch
module. This module is shown in Figure 55.
The address switches are set at the factory to a default GPIB address of
25 (in fact, the binary equivalent of 25).
The GPIB address 25 is also used as the default in the installation of the UV-Visible
N O T E
ChemStation software.
To change the GPIB address of your spectrophotometer:
1 Turn- off your spectrophotometer.
2 On the spectrophotometer's configuration switch module, make sure
Figure 55
Agilent Cary 8454 UV-Visible Spectroscopy System Installation Guide
switches 1 and 2 are in the down position for GPIB address definition
and set the switches 3 through 8 for the appropriate address, see
Table
5.
1
2 3 4
5 6
7 8
Configuration switch module
GPIB Communication, Connections and Configuration
Changing the GPIB Address of the Spectrophotometer
6
109