LAN Communication, Installation, Connection and Configuration
Troubleshooting LAN Communication
Verify that the LAN Interface card is correctly configured
To ensure that all parameters of the LAN interface card are set correctly,
use the Agilent BootP Service program from the
Agilent UV- Visible ChemStation CD- ROM and disable any other means
that configure the LAN interface card of the instrument. Install the bootp
service as described in section
page 120 and configure it for the MAC address of the LAN interface card.
DHCP server
Make sure that no DHCP server runs in the same subnet as the
instrument(s), as these servers also may respond to a bootp request and
do not set the buffer handling of the LAN interface card correctly, i.e.
T145=64:T146=01:T147=01. They also may assign a different IP address to
the instrument each time it is started.
If a DHCP server is running in the same subnet, make sure that fixed IP addresses are
N O T E
reserved for your instruments. Ask you IT consultant to make sure that no IP address
conflict may occur.

Frequent buffer overrun in Agilent Cary 8454 logbook

Please contact your system administrator, to check if the network is not
capable of the network traffic induced due to the data acquisition of the
instrument(s). This can also be caused by an incorrectly configured buffer
handling of the LAN interface card (correct settings:
T145=64:T146=01:T147=01).
134
"Using the Agilent BootP Service"
Agilent Cary 8454 UV-Visible Spectroscopy System Installation Guide