Power Meter Accuracy Test

This test is intended for power meters used in testing the PNA series network analyzers.
The
"Dynamic Accuracy Test"
a higher accuracy than the standard power meter.
Power meters with options G12 and H12 specify an improved instrumentation accuracy
over a limited power range. (These power meters do not contain unique hardware.) A
power meter may be returned to the factory to have one of these options added to an
existing power meter or to renew the calibration for one of these options.
This test procedure is an alternative to returning the power meter to the factory. When a
power meter passes this test, it is considered to be calibrated for the G12 or H12 option
even though it has not been returned to the factory.
This test procedure is available on the Internet as a separate document, Agilent Part
Number E8356-90051. The on-line document may have a more recent print date. Refer to
"Contacting Agilent" on page 2-10
EPM Series Power Meters That Can Be Tested Using This Procedure
This procedure assumes that the recommended model number power meter is being tested.
The alternate model numbers can be tested but the necessary procedural steps may differ.
Recommended
Model Number
E4419B
NOTE
It is recommended that the revision number for the power meter "Main
Firmware" be Ax.03.00 or higher. This applies to all power meter model
numbers listed above (both recommended and alternate).
Equipment Used for the Power Meter Accuracy Test
Equipment Type
Range calibrator
Precision digital voltmeter
Power sensor cable
Cable, BNC, 50Ω
Adapter for connecting BNC
cable to DVM inputs
a. Required accuracy and resolution at the following voltage levels:
14 mV input:
0.140 V input:
0.450 V input:
Service Guide E8364-90026
requires the use of a power meter that has been calibrated to
for the Internet address (URL).
Alternate
Model Numbers
E4418A/B, E4419A, EPM-441A, EPM-442A
Recommended
Model or Part Number
11683A
3458A
8120-8319, 11730A
Any
Any
0.0100% accuracy
0.0050% accuracy
0.003% accuracy
Performance Tests (Agilent N2721A Software Package)
Alternate
Model or Part Number
None
Any with the required accuracy and
resolution
Any equivalent
Any
Any
10 nV resolution
100 nV resolution
100 nV resolution