Invacare Storm Service Manual page 24

Hide thumbs Also See for Storm:
Table of Contents

Advertisement

OTHER
PROBLEM
SYMPTOMS
None
Motors do
not run
Corroded
contacts
None
Motor
makes
clicking
noise
None
Scraping
noise or
motor
blocked
Transmissi
None
on makes
clicking
noise
POSSIBLE
SOLUTION
CAUSE
Bad connections Check all cabling.
Fuse burnt out
Check cabling and
replace fuse
Batteries
Replace batteries
defective
Cabling to power
check cabling
module or
remote defective
Power module
Replace power
defective
module
Water, salt or
Check cabling,
urine has
replace if
penetrated
necessary.
Bearing
Replacing the
defective
motor
Collector
Measure internal
defective
resistance of
motor, replace
motor if defective.
Bearing
Replacing the
defective
motor
Transmission
Replace
defective
transmission.
Transmission
Replace
defective
transmission.
Drive wheel
Tighten drive
loose
wheel, secure
bolts with Loctite if
necessary
Documentation
See chapter 8.3.5.
See chapters 8.3.5 and
8.3.4
See chapter 8.3.2.
See chapter 8.3.5.
See chapter Fehler!
Verweisquelle konnte
nicht gefunden werden..
See chapter 8.3.5.
See chapter 8.17.1.
See chapters 7.1 and
8.17.1
See chapter 8.17.1.
See chapter 8.17.1.
See chapter 8.17.1.
See chapter 8.16.1.
23