Nikon SMZ25 Instruction Manual page 97

Research stereo microscope
Hide thumbs Also See for SMZ25:
Table of Contents

Advertisement

6
Attach the coaxial epi illuminator. (for the EPI System)
Attach the coaxial epi illuminator to the zooming body.
(1)
Completely loosen the M4 set screw for the
illuminator at the top front of the zooming body using
a hex driver (nominal designation: 2).
(2)
Align and fit the round dovetail at the bottom of the
coaxial epi illuminator with the round dovetail of the
zooming body and tighten the fixing screw.
 Attaching the fiber
Loosen the M4 set screws for the fiber on the side of the
two fiber attachment holes on the rear side of the coaxial
epi illuminator using a hex driver (nominal designation: 2).
Insert the fiber to the attachment hole until it reaches the
limit and tighten the fixing screw.
Firmly insert the other end of the fiber into the fiber
attachment hole of the C-FLED2 LED Light Source for
Fiber Illuminator. See the instruction manual for the light
source for details.
 Attaching the 1/4 lambda-plate
Screw and attach the 1/4 lambda plate supplied with the coaxial epi illuminator to the end of the objective.
In case of using a 0.5x objective, remove the protection ring at the end of the objective, and in case of using a 1x
objective, remove the correction glass at the end of the objective before attaching the 1/4 lambda plate.
Chapter 5
Assembling
Attaching the LED fiber by tightening the screw
83
Attaching the coaxial epi illuminator
by tightening the screw