 Attaching a light shielding plate
When using a focus mount, attach the light shielding plate to the zooming body, and when using an intelligent
nosepiece, attach it to the bottom of the intelligent nosepiece.
 Attachment to the zooming body
Fit the ring of the light shielding plate to the periphery of
the dovetail attachment adapter attached to the lower part
of the zooming body, and firmly tighten the M2.5 hexagon
socket head bolt on the rear side of the ring using a hex
driver (nominal designation: 2).
 Attaching to the intelligent nosepiece
(1)
Separate the ring attached to the light shielding
plate. The ring is attached to the light shielding plate
with the two M2.5 hexagon socket head bolts.
Remove them using a hex driver (nominal
designation: 2).
(2)
Align the screw holes of the light shielding plate with
the two screw holes at the bottom front of the
intelligent nosepiece and firmly attach the light
shielding plate using the two M2.5 hexagon socket
head bolts that were used to affix the ring.
Chapter 5
Assembling
Removing the ring from the light shielding plate
86
Attaching the shielding plate
by tightening the screw
Attaching the shielding plate
by tightening the screws