Error Message - Faro FOCUS3D X 130 User Manual

Hide thumbs Also See for FOCUS3D X 130:
Table of Contents

Advertisement

Focus
X 130 Error Messages
3D

Error Message

Internal Error: Invalid
parameter
Distance measurement
tolerance exceeded
WLAN error
Color aquisition failure
Out of time failure
Module Status Error
Scanner Operation Fail-
ure
Command not executed
Description
The scanner parameters are
inconsistent.
The white reference area on
the scanner's base gave
inconsistent measurements.
The WLAN device could
not be found.
Color aquisition has stopped
unexpectedly. Probably
color aquisition is not com-
plete.
This indicates an internal
scanner error.
Module Status Error: Data
version mismatch.
Figure 8-3 Error Message
Internal scanner error.
Command could not be exe-
cuted because of a previous
running scan operation. A
scan is still active, you can-
not start the next operation
now.
Action
Please restart scanner. If the
problem still occurs after
restart, contact FARO Cus-
tomer Service
Please check cleanliness of
this reference area, mirror,
and lenses.
Please restart scanner. If the
problem still occurs after
restart, contact FARO Cus-
tomer Service
Please save log file and con-
tact FARO Customer Service
Please restart scanner. If the
problem still occurs after
restart, contact FARO Cus-
tomer Service
Internal scanner communica-
tion problem. Please ask
FARO Customer Service for
a firmware upgrade.
Please restart scanner. If the
problem still occurs after
restart, contact FARO Cus-
tomer Service
Wait until the scan has been
finished.