Electrical Requirements

General

2.1
■ Power
Problem
There is no power even
though the power switch is
on.
A motorized accessory
does not operate.
Contact arm LED is
blinking.
■ Illumination
Lamp does not light.
■ Controls on the main body/controller
Problem
Brightness cannot be
adjusted with the
dia-illumination brightness
control knob.
The simple remote control
pad does not work to move
the motorized unit.
Chapter 2
Cause
The power cord is not connected, or is
connected improperly.
A contact arm is not used.
The motorized unit connected to the contact arm
is attached incorrectly.
The motorized unit connected to contact arm is
not in correct combination.
The motorized unit connected to contact arm is
not in correct combination.
There is no power supplied.
The lamp has burned out.
The lamp is not attached.
Cause
Operation is disabled.
A contact arm is not used.
Cables for the contact arm, control box, and the
simple remote control pad are not connected
properly.
Troubleshooting
Connect the cable properly.
(→See Chapter 1, "3 Assembly Method, 20 Connect the
motorized device cable" in the "Assembly/Maintenance".)
Use a contact arm.
(→See Chapter 1, "3 Assembly Method, 5 Attach the
standard arm/contact arm" in the
"Assembly/Maintenance".)
Attach the motorized unit correctly.
(→See Chapter 1, "3 Assembly Method" in the
"Assembly/Maintenance".)
Attach the motorized unit in the correct combination.
(→See Chapter 1, "3 Assembly Method - Introduction" in
the "Assembly/Maintenance".)
Attach the motorized unit in the correct combination.
(→See Chapter 1, "3 Assembly Method - Introduction" in
the "Assembly/Maintenance".)
Plug in the power cord.
(→See Chapter 1, "3 Assembly Method, 21 Connect the
power cord" in the "Assembly/Maintenance".)
Replace the lamp with the specified type.
(→See Chapter 3, "1 Replacing the Lamp" in the
"Assembly/Maintenance".)
Attach a designated lamp.
(→See Chapter 1, "3 Assembly Method, 3 Attach the lamp"
in the "Assembly/Maintenance".)
Change the setting.
(→See Chapter 3 "1.1 Adjustment by Lamp Voltage"
-
Turning lamp ON, brightness adjustment control" in
the "Operation".)
Use a contact arm.
(→See Chapter 1, "3 Assembly Method - 5 Attach the
standard arm/contact arm" in the
"Assembly/Maintenance".)
Connect the cable properly.
(→See Chapter 1, "3 Assembly Method - 20 Connect the
motorized device cable" in the "Assembly/Maintenance".)