Electrical - Nikon ECLIPSE TS100 Instructions Manual

Inverted microscope
Hide thumbs Also See for ECLIPSE TS100:
Table of Contents

Advertisement

Electrical

3
Lamp does not turn on even when power switch is on.
No electricity.
Power cord is not connected to the
microscope main body.
Lamp is not installed.
(for TS100/TS100-F only)
Lamp is burnt out.
(for TS100/TS100-F only)
Not using the specified lamp.
(for TS100/TS100-F only)
Fuses are blown.
(for TS100/TS100-F only)
Lamp cord is disconnected.
Lamp is about to burn out.
(for TS100/TS100-F only)
Bad connection.
Fuse holder is not attached securely.
(for TS100/TS100-F only)
Room voltage is fluctuating erratically.
(for TS100/TS100-F only)
Lamp is not securely inserted in socket.
(for TS100/TS100-F only)
6 Troubleshooting
3 Electrical
Causes
Lamp flickers. Brightness is unstable.
Causes
E-44
Corrective Measures
Insert the power cord into an electrical
outlet. (P.35)
Insert the power cord into the AC inlet.
(P.35)
Install the lamp. (P.34)
Replace the lamp. (P.34)
Use the specified lamp. (P.34)
Replace the fuses. (Use only the
specified type.) (P.34)
Connect to connector on rear of
microscope. (P.34)
Corrective Measures
Replace the lamp. (P.34)
Secure all connections.
Attach the holder securely. (P.34)
Use an AC power supply with stable
current meeting local safety regulations.
Insert the lamp securely. (P.34)